FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240863 · Received November 25, 2008

Report

Report Number
6000001-2007-04119
Event Type
Malfunction
Date Received
November 25, 2008
Date of Event
February 3, 2007
Report Date
February 3, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE CONDITION OF FAIL CODE 814:01 WAS OBSERVED IN THE EVENT HISTORY. THIS FAIL CODE WAS CAUSED BY DEPLETED BATTERIES. THE BATTERIES WERE REPLACED. THE ISSUE OF FAIL CODE 814:01 IS CURRENTLY BEING INVESTIGATED UNDER CAPA. THE ISSUE OF DEPLETED BATTERIES IS BEING INVESTIGATED UNDER CAPA. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE.

Description of Event or Problem · 1

THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN NOTED FAIL CODE 814:01 IN THE EVENT HISTORY. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA