COLLEAGUE CX VOLUMETRIC INFUSION PUMP
MDR report key: 1240852
·
Received November 25, 2008
Report
- Report Number
- 6000001-2007-04118
- Event Type
- Malfunction
- Date Received
- November 25, 2008
- Date of Event
- January 1, 2007
- Report Date
- January 15, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE CONDITION OF FAIL CODE 570 WAS CONFIRMED. THIS FAIL CODE WAS CAUSED BY DEPLETED BATTERIES. THE BATTERIES WERE REPLACED. THIS ISSUE IS CURRENTLY BEING INVESTIGATED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED AN INFUSION PUMP WITH FAILURE CODE 570 DURING BIOMED TESTING. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.