BD ASEPTO¿ SCALP INFUSION SET
Report
- Report Number
- 9610048-2021-00122
- Event Type
- Malfunction
- Date Received
- September 1, 2021
- Date of Event
- August 4, 2021
- Report Date
- October 19, 2021
- Manufacturer
- BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA.
- Product Code
- FPA
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: A PHYSICAL SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BUT BD WAS PROVIDED WITH A PHOTO OF THE ISSUE FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED FOR THE REPORTED LOT, 1032509, AND NO QUALITY ISSUES WERE FOUND DURING PRODUCTION. OUR QUALITY ENGINEER REVIEWED THE PROVIDED PHOTO AND OBSERVED THAT THE TUBE WAS CUT AND STICKING OUT FROM THE EDGE OF THE PACKAGING. BASED OFF THE PROVIDED PHOTO THE ENGINEER WAS ABLE TO VERIFY THE REPORTED DEFECT. IT WAS DETERMINED THAT THIS WAS AN OPERATOR ERROR THAT OCCURRED DURING THE PACKAGING PROCESS. H3 OTHER TEXT : SEE H10.
IT WAS REPORTED THAT THE BD ASEPTO¿ SCALP INFUSION SET PUNCTURED THROUGH THE PACKAGING UNIT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM PORTUGUESE TO ENGLISH: "EXTENSION OF THE MATERIAL SEALED WITH THE PACKAGING, BREAKING IT."
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD ASEPTO" SCALP INFUSION SET PUNCTURED THROUGH THE PACKAGING UNIT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "EXTENSION OF THE MATERIAL SEALED WITH THE PACKAGING, BREAKING IT."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1301892 | BD ASEPTO¿ SCALP INFUSION SET | INTRAVENOUS INFUSION SET | FPA | BECTON DICKINSON INDUSTRIAS CIRURGICAS, LTDA. | 1032509 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |