FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240666 · Received November 25, 2008

Report

Report Number
6000001-2007-04012
Event Type
Malfunction
Date Received
November 25, 2008
Date of Event
January 15, 2007
Report Date
January 15, 2007
Manufacturer
BAXTER HEALTHCARE PTE.LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: DEVICE WAS EVALUATED ON-SITE AND THE REPORTED CONDITION OF FAILURE CODE 570 WAS CONFIRMED DUE TO DEPLETED BATTERIES. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES WITH 4 DISCHARGES BELOW ALARM THRESHOLD CAUSED FAILURE CODE 570. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPORTED AN INFUSION PUMP WITH A FAILURE CODE 570. THE EVENT WAS REPORTED TO HAVE OCCURRED DURING BIO-MED TESTING. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE.LTD. N/A NA