FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240624 · Received November 25, 2008

Report

Report Number
6000001-2007-04096
Event Type
Malfunction
Date Received
November 25, 2008
Date of Event
January 1, 2007
Report Date
January 15, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: EVALUATION OCCURRED ON-SITE. THE CONDITION OF DEPLETED BATTERIES WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE PUMP'S BATTERIES WERE POTENTIALLY DAMAGED AND WERE THEREFORE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

A BAXTER FIELD SERVICE ENGINEER REPORTED AN INFUSION PUMP WITH DEPLETED BATTERIES. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE DEVICE WAS IN PATIENT USE WHEN THE EVENT OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. N/A NA

Patients

Seq Age Sex Outcome Treatment
1