FDA Adverse Event Malfunction Summary report: N

COLLEAGUE VOLUMETRIC INFUSION PUMP

MDR report key: 1240410 · Received November 25, 2008

Report

Report Number
6000001-2007-06112
Event Type
Malfunction
Date Received
November 25, 2008
Date of Event
April 1, 2007
Report Date
April 4, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
2005-047-MD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY REPORTED AN INFUSION PUMP THAT READS FAILURE. EVALUATION SUMMARY: DURING PRODUCT EVALUATION, A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS OBSERVED. FAILURE CODE 703:00, 403:317:864:000, AND 402:317:858:000 IN THE EVENT HISTORY CONFIRMS THE DEFECTIVE UIM PCB. THESE FAILURE CODES ARE MANIFESTED AS A RESULT OF THE UIM PCB BEING DEFECTIVE. THE UIM PCB WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO BAXTER. DURING TESTING, THE BAXTER TECHNICIAN REPORTED AN INFUSION PUMP WITH A DEFECTIVE UIM (USER INTERFACE MODULE) PCB (PRINTED CIRCUIT BOARD). ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAS BEEN REPORTED RELATED TO THIS DEVICE. NO ADDITIONAL INFORMATION IS KNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA