FDA Adverse Event Other Summary report: N

FLUOROCON

MDR report key: 12404 · Received March 30, 1994

Report

Report Number
MW1001326
Event Type
Other
Date Received
March 30, 1994
Report Date
March 28, 1994
Manufacturer
PILKINGTON BARNES HIND, INC.
Product Code
HPX
Report Source
Voluntary report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THIS LENS WAS NEVER DISPOSED TO THE PT. BASE CURVE VERIFICATION WAS 8/12 INSTEAD OF 8/07, WELL BEYOND ACCEPTED TOLERANCE OF GOOD MFG PRACTICES AND ANSI STANDARDS. THIS WAS SUPPOSED TO HAVE BEEN RE-INSPECTED FOR ACCURACY BEFORE SHIPMENT. CORNEAL HEALTH COULD HAVE BEEN SERIOUSLY COMPROMISED BY THIS POOR QUALITY CONTROL

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLUOROCON RIGID GAS PERMEABLE CONTACT LENS HPX PILKINGTON BARNES HIND, INC. 8.07/9.5 DIAM ORDER #4898938/001

Patients

Seq Age Sex Outcome Treatment
1 * Other