FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240373 · Received November 25, 2008

Report

Report Number
6000001-2007-04161
Event Type
Malfunction
Date Received
November 25, 2008
Date of Event
January 1, 2007
Report Date
January 12, 2007
Manufacturer
BAXTER HEALTHCARE PTE.LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: EVALUATION WAS PERFORMED AND THE CONDITION OF FAILURE CODE 500:320:844:0000 WAS CONFIRMED. INSPECTION OF THE PUMP REVEALED FAILURE CODE 500:320:844:0000 WAS CAUSED BY PUMP HEAD MODULE DUE TO INTERMITTENT FAILURE. THE PUMP HEAD MODULE WAS REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPORTED AN INFUSION PUMP WITH A FAILURE CODE 500:320:844:0000. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT HE DEVICE WAS IN PATIENT USE WHEN THE EVENT OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE.LTD. N/A NA

Patients

Seq Age Sex Outcome Treatment
1