FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-04722
- Event Type
- Malfunction
- Date Received
- November 24, 2008
- Date of Event
- January 1, 2007
- Report Date
- January 11, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE FACILITY REPRESENTATIVE REPORTED AN UNSPECIFIED ALARM SITUATION. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURRED DURING A PATIENT INFUSION.EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. INSPECTION OF THE DEVICE FOUND THAT A CONSTANT DOWNSTREAM OCCLUSION ALARM WAS CAUSED BY A BROKEN DOOR ASSEMBLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, A BROKEN DOOR ASSEMBLY WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLO-GARD 6201 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE (SG) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |