FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240286 · Received November 24, 2008

Report

Report Number
6000001-2007-03973
Event Type
Malfunction
Date Received
November 24, 2008
Date of Event
January 1, 2007
Report Date
January 11, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED CONDITION OF FAILURE CODE 812:02 ON CHANNEL A WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. FAILURE CODE 812:02 WAS CAUSED BY A FAULTY PUMP HEAD MECHANISM. PUMP HEAD MECHANISM CHANNEL A WAS THEREFORE REPLACED. DEVICE WAS EVALUATED IN 2007.

Description of Event or Problem · 1

THE FIELD SERVICE ENGINEER REPORTED A PUMP WITH FAILURE CODE 812:02 ON CHANNEL A. THIS PROBLEM WAS IDENTIFIED DURING BIO-MED TESTING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1