INFUSOR LV 5 ML/H SYSTEM
Report
- Report Number
- 6000001-2007-06045
- Event Type
- Malfunction
- Date Received
- November 24, 2008
- Date of Event
- March 30, 2007
- Report Date
- April 9, 2007
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: ONE USED UNIT WAS RECEIVED CONTAINING NO SOLUTION. UPON RECEIPT, THE UNIT WAS VISUALLY EXAMINED FOR SIGNS OF ABNORMALITY; HOWEVER, NONE WERE FOUND. THE IMPRINT ON THE FLOW RESTRICTOR WAS NOTED TO BE CORRECT. PER PROCEDURE, A FLOW TEST WAS SUBSEQUENTLY PERFORMED ON THE UNIT FOR 43.5 HOURS. AT THE END OF THE FLOW TEST PERIOD, THE FOLLOWING FLOW RATE (ML/HR) WAS PRODUCED FROM THE UNIT : CALCULATED 5.05 NORMALIZED (2.5%) 5.18 SPECIFICATION RANGE 4.50 ---5.50. THE FLOW RATE WAS FOUND TO BE WITHIN THE SPECIFICATION RANGE. THEREFORE, BASED ON THE EVALUATION FINDINGS, THE DEVICE PERFORMED AS EXPECTED. A REVIEW OF ALL RECORDS RELATED TO THE MANUFACTURE OF LOT 06M025 HAS BEEN CONDUCTED, WHICH REVEALED NO EVIDENCE OF RELATED DEFECTS DURING INSPECTION, TESTING AND MANUFACTURING OF THE PRODUCT LOT. THE MANUFACTURING FACILITY HAS BEEN MADE AWARE OF THIS REPORT THROUGH BAXTER'S COMPLAINT HANDLING SYSTEM. THE MANUFACTURING FACILITY WILL CONTINUE TO MONITOR SIMILAR INCIDENTS FOR POSSIBLE TRENDS.
NATIONAL COMPLAINT COORDINATOR FOR BAXTER JAPAN REPORTED AN ALLEGED OVERINFUSION THAT OCCURRED ON ONE INFUSOR DEVICE DURING PATIENT INFUSION VIA AN INTRAVENOUS INJECTION. THE DEVICE WAS FILLED WITH 30ML OF 5-FU AND 195ML OF NORMAL SALINE FOR TOTAL FILL VOLUME OF 225ML. THE EXPECTED INFUSION TIME WAS 46 HOURS. THE ACTUAL INFUSION WAS UNKNOWN. THE DEVICE WAS NOT USED PRIOR TO THE INCIDENT. THERE WERE NO REPORTS OF INJURY OR MEDICAL INTERVENTION RELATED TO THE REPORTED CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR LV 5 ML/H SYSTEM | ELASTOMERIC INFUSION SYSTEM | MEB | BAXTER HEALTHCARE CORPORATION | NA | 06M025 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |