FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
MDR report key: 1240232
·
Received November 24, 2008
Report
- Report Number
- 6000001-2007-06173
- Event Type
- Malfunction
- Date Received
- November 24, 2008
- Date of Event
- April 1, 2007
- Report Date
- April 12, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 2005-047-MD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE REPORTED CONDITION OF FAILURE CODE 703 WAS CONFIRMED IN THE PUMP'S EVENT HISTORY FILE DURING PRODUCT EVALUATION. DURING PRODUCT EVALUATION, A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS OBSERVED. THE REPORTED CONDITION OF FAILURE CODE 703 CONFIRMS THE DEFECTIVE UIM PCB. THE UIM PCB WAS REPLACED AND THE DEVICE WAS TESTED.
Description of Event or Problem · 1
THE FACILITY REPORTED A PUMP WITH FAILURE CODE 703. THIS PROBLEM WAS IDENTIFIED AFTER USE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.