FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240221 · Received November 24, 2008

Report

Report Number
6000001-2007-05884
Event Type
Malfunction
Date Received
November 24, 2008
Date of Event
April 1, 2007
Report Date
April 12, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY REPRESENTATIVE REPORTED FAILURE CODE 814:04. EVALUATION SUMMARY: DURING PRODUCT EVALUATION, THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) VALUES WERE FOUND TO BE OUT OF SPECIFICATION. THE AIL PCB WAS REPLACED. THE REPORTED CONDITION OF A DEFECTIVE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) WILL BE CLOSED.

Description of Event or Problem · 1

THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING SERVICE THE BAXTER REPAIR TECHNICIAN NOTED A FAULTY AIR IN LIN PRINTED CIRCUIT BOARD. THE HOSPITAL REPRESENTATIVE STATED THAT THERE HAVE BEEN NO REPORTS OF ANY PATIENT INCIDENT INVOLVING THIS PUMP. NO ADDITIONAL INFORMATION IS AVAILABLE.