FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240173 · Received November 24, 2008

Report

Report Number
6000001-2007-05957
Event Type
Malfunction
Date Received
November 24, 2008
Date of Event
April 13, 2007
Report Date
April 13, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: DURING PRODUCT EVALUATION, AN OUT OF SPECIFICATION AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) WAS CONFIRMED. THE AIL PCB WAS REPLACED AND THE DEVICE WAS TESTED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION, AN OUT OF SPECIFICATION AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) WAS OBSERVED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION NECESSARY ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1