FDA Adverse Event
Malfunction
Summary report: N
FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
MDR report key: 1240168
·
Received November 24, 2008
Report
- Report Number
- 6000001-2007-06135
- Event Type
- Malfunction
- Date Received
- November 24, 2008
- Date of Event
- April 1, 2007
- Report Date
- April 10, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION RESULTS: THE REPORTED CONDITION OF THE OVERINFUSION WAS NOT DUPLICATED OR CONFIRMED. THE DEVICE PASSED ALL VISUAL AND FUNCTIONAL TESTS PRIOR TO CUSTOMER RETURN.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED, "INFUSING TOO FAST." INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE PROBLEM OCCURRED DURING A PATIENT INFUSION. ALTHOUGH BAXTER ATTEMPTED TO OBTAIN INFORMATION, DETAILS WERE NOT AVAILABLE REGARDING ADDITIONAL CONTACT INFORMATION. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLO-GARD 6201 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE (SG) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|