FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP

MDR report key: 1240109 · Received November 24, 2008

Report

Report Number
6000001-2007-06125
Event Type
Malfunction
Date Received
November 24, 2008
Date of Event
April 1, 2007
Report Date
April 10, 2007
Manufacturer
BAXTER HEALTHCARE PTE.LTD.
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: DURING PRODUCT EVALUATION, REPORTED CONDITION OF WILL NOT TURN ON AND PROBLEMS WITH LOAD TUBING WAS CONFIRMED. THE PUMP WAS RECEIVED WITH CHANNEL A AND CHANNEL B OPEN. THE PUMP WILL NOT POWER ON WITH AN OPEN CHANNEL. FAILURE CODE 715:00 ON CHANNEL A WAS FOUND IN THE EVENT HISTORY. FAILURE CODE 715:00 HAVE KNOWN TO OCCUR WHEN THE MANUAL TUBE RELEASE IS TIMED-OUT OR USED IMPROPERLY MULTIPLE TIMES. CHANNEL A AND CHANNEL B WERE RESET USING MANUAL TUBE RELEASE KNOBS. THE PUMP PASSED POWER ON SELF-TEST ON AC AND BATTERIES WITHOUT ANY FAILURE. DURING EVALUATION FROM THE EVENT HISTORY IT WAS CONFIRMED THAT THIS FAILURE WAS NOT OCCURRED DURING INFUSION. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.

Description of Event or Problem · 1

THE FACILITY REPORTED AN INFUSION PUMP THAT WILL NOT TURN ON AND HAS TUBE LOADING PROBLEMS. THE EVENT WAS REPORTED TO HAVE OCCURRED DURING PATIENT USE. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE.LTD. N/A NA

Patients

Seq Age Sex Outcome Treatment
1