FDA Adverse Event
Malfunction
Summary report: N
FLO-GARD 6201 VOLUMETRIC INFUSION PUMP
MDR report key: 1240074
·
Received November 24, 2008
Report
- Report Number
- 6000001-2007-06276
- Event Type
- Malfunction
- Date Received
- November 24, 2008
- Date of Event
- April 1, 2007
- Report Date
- April 6, 2007
- Manufacturer
- BAXTER HEALTHCARE (SG)
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FACILITY REPRESENTATIVE REPORTED AN OCCLUSION ALARM FOUND DURING BIO-MED TESTING. EVALUATION SUMMARY: THE PUMP WAS EVALUATED BY A BAXTER SERVICE TECHNICIAN. INSPECTION OF THE DEVICE FOUND THAT THAT OCCLUSION ALARM WAS CAUSED BY A BROKEN DOOR ASSEMBLY (DOOR AND DOOR LATCH.) REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE BY BAXTER, A BROKEN DOOR WITH A BROKEN DOOR LATCH WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLO-GARD 6201 VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE (SG) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |