FDA Adverse Event Malfunction Summary report: N

PCA LL PUMP

MDR report key: 1240069 · Received November 24, 2008

Report

Report Number
6000001-2007-05991
Event Type
Malfunction
Date Received
November 24, 2008
Date of Event
April 1, 2007
Report Date
April 19, 2007
Manufacturer
BAXTER HEALTHCARE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BATTERY LEAKING WAS FOUND DURING SERVICE. CORROSION WAS FOUND ON THE MECHANICAL BOARD AND MPU BOARD. PUMP WAS RETURNED UNREPAIRED PER THE CUSTOMERS REQUEST.

Description of Event or Problem · 1

DURING SERVICE BATTERY LEAKAGE WAS FOUND ON PUMP. BATTERY CORROSION WAS FOUND ON MPU BOARD AND MECHANICAL BOARD. BAXTER TECHNICIAN RECOMMENDED CUSTOMER USE RECHARGEABLE BATTERIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PCA LL PUMP ANESTHESIA PUMPS FRN BAXTER HEALTHCARE N/A NA

Patients

Seq Age Sex Outcome Treatment
1