ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-02411
- Event Type
- Malfunction
- Date Received
- August 30, 2021
- Report Date
- December 10, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H10: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1033145 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 190-000 / LOT 1033145. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING), RELATED TO KIT LOT 1033145 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS PATIENT SAMPLE INTERFERENCE.
REFERENCE MFR. REPORT: 1221359-2021-02397, 1221359-2021- 02399, 1221359-2021-02409, 1221359-2021-02412. THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CUSTOMER REPORTED SEVEN FALSE NEGATIVE RESULTS WITH THE ID NOW COVID-19 ASSAY ON MULTIPLE LOT NUMBERS PERFORMED ON VARIOUS DATES. THIS MFR. REPORT ADDRESSES FOUR (4) OF FIVE (5). THE CUSTOMER REPORTED FALSE NEGATIVE RESULT ON THE ID NOW COVID-19 ASSAYS PERFORMED ON (B)(6) 2021 ON A NASAL SWAB. ADDITIONAL TESTING ON A NASOPHARYNGEAL (NP) IN VIRAL TRANSPORT MEDIA PERFORMED ON (B)(6) 2021 WITH BIOFIRE PLATFORM (RESULT DETECTED). NO ADDITIONAL PATIENT INFORMATION, INCLUDING SYMPTOMS, TREATMENT, IMPACT AND OUTCOME WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1286465 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 1033145 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |