FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 12387811 · Received August 30, 2021

Report

Report Number
1221359-2021-02469
Event Type
Malfunction
Date Received
August 30, 2021
Report Date
December 3, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M150761 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: M150761 TEST BASE PART NUMBER 190-430 / LOT: M150761. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M150761 SHOWED THAT THE COMPLAINT RATE IS (B)(4).. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE AN ASSIGNABLE ROOT CAUSE AS THE LOG FILES WERE NOT PROVIDED FOR INVESTIGATION. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED AN UNCONFIRMED FALSE NEGATIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON A THROAT AND NARES SAMPLE. ALTHOUGH REQUESTED, NO ADDITIONAL INFORMATION HAS BEEN PROVIDED AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1280418 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. M150761

Patients

Seq Age Sex Outcome Treatment
1 Unknown