CORTRAK2 NG/NI FEEDING TUBE W/ELECTROMAG TRANSMITTING STYLET W/ ENFIT CONNECTOR
Report
- Report Number
- 9611594-2021-00115
- Event Type
- Malfunction
- Date Received
- August 30, 2021
- Date of Event
- August 8, 2021
- Report Date
- September 24, 2021
- Manufacturer
- AVANOS MEDICAL INC.
- Product Code
- KNT
- UDI-DI
- 00350770460475
- PMA / PMN Number
- K821906
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE PRODUCT INVOLVED IN THE REPORT HAS BEEN RETURNED AND IS BEING PROCESSED FOR EVALUATION. ALL INFORMATION REASONABLY KNOWN AS OF 13 SEP 2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT COMP-GHC-21-02508. THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
ONE USED SAMPLE WAS RETURNED FOR EVALUATION. VISUAL EXAMINATION OF THE SAMPLE REVEALED DISCOLORATION AND DAMAGE. DURING DECONTAMINATION, A BROWNISH SUBSTANCE WAS FLUSHED FROM THE TUBING. A SPLIT/TEAR WAS NOTED AT THE 9CM MARKING. THE TUBING HAD EXPANDED INTO A BALLOON SHAPE, WHICH THEN BURST, CAUSING A SEPARATION OF THE TUBE INTO TWO PIECES. THE DISTAL END OF THE TUBE WAS NOT RETURNED. THE COMPLAINT IS CONFIRMED AS REPORTED. ROOT CAUSE COULD NOT BE CONCLUSIVELY DETERMINED, HOWEVER, IT IS LIKELY A USER RELATED PROBLEM SINCE AS PER THE INSTRUCTIONS FOR USE, VIGOROUS SYRINGE FORCE SHOULD NOT BE USED TO IRRIGATE, ADMINISTER LIQUIDS, OR UNBLOCK THE TUBE. ALL INFORMATION REASONABLY KNOWN AS OF 23 SEP 2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT COMP-(B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
PER ADDITIONAL INFORMATION RECEIVED 8 SEP 2021, THE FRAGMENT PASSED ON ITS OWN BUT WAS NOT RECOVERED. THE PATIENT IS "DOING FINE."
THE SAMPLE IS REPORTED TO BE AVAILABLE, BUT HAS NOT YET BEEN RECEIVED BY THE MANUFACTURER. A REVIEW OF THE DEVICE HISTORY RECORD IS NOT POSSIBLE AS NO LOT NUMBER WAS PROVIDED. ALL INFORMATION REASONABLY KNOWN AS OF 27 AUG 2021 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY AVANOS MEDICAL INC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO AVANOS MEDICAL INC.. AVANOS MEDICAL INC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE AVANOS MEDICAL COMPLAINT DATABASE AND IDENTIFIED AS COMPLAINT (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN AVANOS MEDICAL INC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.
IT WAS REPORTED THAT 10CM OF THE DISTAL END OF THE NASALGASTRIC (NG) TUBE BROKE OFF INSIDE THE PATIENT. PER X-RAY, THE PIECE IS INSIDE THE PATIENT'S INTESTINES AND THEY ARE WAITING FOR THE PIECE TO PASS NATURALLY. NO HARM TO THE PATIENT. PER ADDITIONAL INFORMATION RECEIVED ON 20 AUG 2021, THE TUBE WAS IN PLACE FOR 38 DAYS. THE TUBE WAS PLACED USING THE CORTRAK2 MONITOR SN CORTRAK MONITOR. NO FURTHER INFORMATION PROVIDED AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1280812 | CORTRAK2 NG/NI FEEDING TUBE W/ELECTROMAG TRANSMITTING STYLET W/ ENFIT CONNECTOR | DH CORTRAK DISPOSABLES | KNT | AVANOS MEDICAL INC. | 40-9431TRAK2 | UNKNOWN | 00350770460475 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | CORTRAK2 MONITOR SN CORTRAK MONITOR |