FDA Adverse Event Injury Summary report: N

PITT EASY ABUTMENT

MDR report key: 1238223 · Received November 21, 2008

Report

Report Number
9611820-2008-00001
Event Type
Injury
Date Received
November 21, 2008
Report Date
October 29, 2008
Manufacturer
SYBRON IMPLANT SOLUTIONS GMBH
Product Code
NHA
PMA / PMN Number
K053242
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE DOCTOR REPORTED THAT THE PATIENT IS DOING FINE. THE DOCTOR HAD THE PROSTHESIS PLACED IN 2007. IN 2008, THE PATIENT WENT BACK TO SEE THE DOCTOR BECAUSE OF THE FRACTURED ABUTMENT. DURING THIS 1 ½ YEAR TIME PERIOD, THE PATIENT HAD NEVER HAD FOLLOW-UP APPOINTMENTS WITH THE DOCTOR. THE DOCTOR EVALUATED THAT A REBASE OF THE PROSTHESES SHOULD HAVE TAKEN PLACE MUCH EARLIER BECAUSE HE NOTICED THAT THE ABUTMENT WAS OVERLOADED THE ENTIRE TIME. THE PRODUCT WAS RETURNED TO SYBRON IMPLANT SOLUTIONS GMBH FOR EVALUATION. VISUAL EXAMINATION OF THE RETURNED PRODUCT INDICATED THAT THERE WAS NO MATERIAL FAILURE; PLEASE SEE THE ATTACHED MICROSCOPIC PICTURES. THE CAUSE OF THE ABUTMENT FRACTURE WAS OVERLOADING. THIS IS THE FINAL REPORT.

Description of Event or Problem · 1

IN 2008, A DOCTOR REPORTED TO COMPANY THAT A PITT EASY IMPLANT ABUTMENT HAD FRACTURED DURING USE IN A PATIENT'S MOUTH, BUT NOTHING WAS SWALLOWED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PITT EASY ABUTMENT ABUTMENT, IMPLANT, DENTAL, ENDOSSEOUS NHA SYBRON IMPLANT SOLUTIONS GMBH 40051 706118D1

Patients

Seq Age Sex Outcome Treatment
1 Other