FDA Adverse Event Malfunction Summary report: N

PLATE COLUMBIA AG 5PRCT SB 90MM

MDR report key: 12381186 · Received August 28, 2021

Report

Report Number
9680577-2021-00053
Event Type
Malfunction
Date Received
August 28, 2021
Date of Event
August 11, 2021
Report Date
October 22, 2021
Manufacturer
BECTON DICKINSON GMBH
Product Code
KZI
PMA / PMN Number
PREAMENDMENT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: THIS MEMO IS TO SUMMARIZE FINDINGS ON THE RECENT COMPLAINT AGAINST COLUMBIA AGAR WITH 5% SHEEP BLOOD, CATALOG NUMBER 254071, LOT NUMBER 1173789. IT WAS REPORTED THAT 40 PLATES WOULD BE CONTAMINATED. THE COMPLAINT TRENDS WERE REVIEWED. THERE WAS ONE SIMILAR CONFIRMED COMPLAINT REPORTED DURING THAT PERIOD ON THIS LOT NUMBER. THEREFORE, A TREND COULD NOT BE IDENTIFIED. THE BATCH HISTORY REVIEW DID NOT INDICATE ANY DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY, AND NO DEVIATIONS WERE OBSERVED. A SAMPLE WAS NOT PROVIDED FOR FURTHER EVALUATION. A VISUAL INSPECTION WAS PERFORMED ON THE RETENTION SAMPLE AND ONE STACK OF PLATES WERE INCUBATED. AS A RESULT OF THIS ANALYSIS NO CONTAMINATION WAS OBSERVED. BASED ON THE EVALUATION OF THE PROVIDED REPORT AND AS NO PICTURES WERE PROVIDED, THE COMPLAINT WAS NOT CONFIRMED. A CORRECTIVE AND PREVENTIVE ACTION WILL NOT BE IMPLEMENTED AS A TREND COULD NOT BE IDENTIFIED. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING 40 PLATE COLUMBIA AG 5PRCT SB 90MM CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "40 X CONTAMINATED PLATES"

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING 40 PLATE COLUMBIA AG 5PRCT SB 90MM CONTAMINATION WAS OBSERVED BY THE LABORATORY PERSONNEL. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "40 X CONTAMINATED PLATES".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1278603 PLATE COLUMBIA AG 5PRCT SB 90MM CULTURE MEDIA, ENRICHED KZI BECTON DICKINSON GMBH 1173789

Patients

Seq Age Sex Outcome Treatment
1 Unknown