FDA Adverse Event Malfunction Summary report: N

SYRINGE S2 5ML 22GA 1-1/4IN BD CHINA

MDR report key: 12379049 · Received August 27, 2021

Report

Report Number
3002682307-2021-00444
Event Type
Malfunction
Date Received
August 27, 2021
Date of Event
July 29, 2021
Report Date
November 2, 2021
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 2010170. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. TO AID IN THE INVESTIGATION OF THIS ISSUE, A PICTURE OF THE AFFECTED SAMPLES WAS RETURNED FOR EVALUATION BY OUR QUALITY ENGINEER TEAM. THROUGH EXAMINATION OF THE PICTURE, IT APPEARED THAT THE CUTTING OF THE BLISTER PACKAGES HAD BEEN MISALIGNED, CAUSING THE BLISTER TO BE OPEN ON ONE SIDE. THIS TYPE OF DEFECT RESULTED IN THE PRIMARY PACKAGING MACHINE. THE MACHINE HAS A CUTTING SYSTEM AND DUE TO A PUNCTUAL FAILURE, INCORRECT CUTS WERE MADE. THE CUTTING OF THE BLISTER PACKAGES MUST BE DONE CORRECTLY TO KEEP THE SEALED AREA OF EACH UNIT. SINCE THE BLISTER CUTTING SYSTEM WAS MISALIGNED, THE SEALED AREA WAS NOT MAINTAINED WHICH CAUSED THE BLISTER TO OPEN. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT 6 SYRINGE S2 5ML 22GA 1-1/4IN BD CHINA HAD A DAMAGED PACKAGE THAT COMPROMISED STERILITY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "6 SYRINGES, CLOSE TO THE LOWER END WHEN SEALING THE EDGE, NOT SEALED."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 6 SYRINGE S2 5ML 22GA 1-1/4IN BD CHINA HAD A DAMAGED PACKAGE THAT COMPROMISED STERILITY. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "6 SYRINGES, CLOSE TO THE LOWER END WHEN SEALING THE EDGE, NOT SEALED".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1276595 SYRINGE S2 5ML 22GA 1-1/4IN BD CHINA PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2010170

Patients

Seq Age Sex Outcome Treatment
1 Unknown