FDA Adverse Event Malfunction Summary report: N

CONSTELLATION VISION SYSTEM

MDR report key: 12375063 · Received August 26, 2021

Report

Report Number
2028159-2021-00949
Event Type
Malfunction
Date Received
August 26, 2021
Report Date
October 26, 2021
Manufacturer
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
Product Code
HQC
UDI-DI
00380657530410
PMA / PMN Number
K101285
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CORRECTED INFORMATION HAS BEEN PROVIDED IN SECTION B.5 AND E.1. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

THE COMPANY REPRESENTATIVE PROVIDED A PHONE FIX FOR THE CUSTOMER. THE ISSUE WAS CAUSED BY USER ERROR. THE COMPANY REPRESENTATIVE DISCOVERED THAT THE NEW TECH AT THE CUSTOMER SITE DID NOT KNOW HOW TO USE THE SYSTEM CORRECTLY. THE SYSTEM MANUFACTURING DEVICE HISTORY RECORD (DHR) WAS REVIEWED. BASED ON QA ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO USER ERROR. MANUFACTURER WILL CONTINUE TO MONITOR DATA FOR EVIDENCE OF ADVERSE TRENDING AND TAKE FURTHER ACTION, AS APPROPRIATE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

ADDITIONAL INFORMATION PREVIOUSLY REPORTED INDICATED THAT THE OPERATING ROOM TECHNICIAN WAS NEW AND DID NOT OPERATE THE CONSOLE CORRECTLY.

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 1

A HEALTH PROFESSIONAL REPORTED THAT TABLE TOP BULB DID NOT ILLUMINATE ON OPHTHALMIC OPERATING CONSOLE. PROCEDURE DETAILS AND PATIENT IMPACT HAVE NOT BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1272179 CONSTELLATION VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER NA NA 00380657530410

Patients

Seq Age Sex Outcome Treatment
1