FDA Adverse Event Malfunction Summary report: N

INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE

MDR report key: 12374672 · Received August 26, 2021

Report

Report Number
1037905-2021-00437
Event Type
Malfunction
Date Received
August 26, 2021
Report Date
September 22, 2021
Manufacturer
COOK ENDOSCOPY
Product Code
PKL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED QUANTITY HAS BEEN UPDATED FROM TWO (2) TO FIVE (5). INITIAL REPORTER OCCUPATION: PURCHASING, FACILITIES, & IT COORDINATOR. PMA/510(K) # K192697. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF DEPLOYMENT DIFFICULTY FOR INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 0

DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED FIVE (5) COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE CLIPS ARE HARD TO DEPLOY/DETACH. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Additional Manufacturer Narrative · 1

INITIAL REPORTER OCCUPATION: PURCHASING, FACILITIES, & IT COORDINATOR. PMA/510(K) # K192697. INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD COULD NOT BE CONDUCTED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. INVESTIGATION CONCLUSION: A CORRECTIVE ACTION HAS BEEN INITIATED TO REDUCE OCCURRENCES OF DEPLOYMENT DIFFICULTY FOR INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE PRODUCT SAID TO BE INVOLVED IS INCLUDED IN THE SCOPE OF THE CORRECTIVE ACTIONS. PRIOR TO DISTRIBUTION, ALL INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC PROCEDURE, THE PHYSICIAN USED TWO (2) COOK INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICES. THE CLIPS ARE HARD TO DEPLOY/DETACH. AN UNINTENDED SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1270697 INSTINCT PLUS ENDOSCOPIC CLIPPING DEVICE PKL, LIGATOR, HEMORRHOIDAL PKL COOK ENDOSCOPY UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS ENDOSCOPE, PCFHQ190L 2101725