FDA Adverse Event Malfunction Summary report: N

BD RAPID DETECTION OF SARS-COV-2 VERITOR

MDR report key: 12373184 · Received August 26, 2021

Report

Report Number
3006948883-2021-00895
Event Type
Malfunction
Date Received
August 26, 2021
Date of Event
August 8, 2021
Report Date
September 21, 2021
Manufacturer
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
Product Code
QKP
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY: THIS STATEMENT IS TO SUMMARIZE THE INVESTIGATION RESULTS REGARDING THE COMPLAINT THAT ALLEGES FALSE POSITIVE RESULTS WHEN USING KIT RAPID DETECTION OF SARS-COV-2 VERITOR CE (MATERIAL # 256089), BATCH NUMBER 1181838. BD QUALITY PERFORMS A SYSTEMATIC APPROACH TO INVESTIGATE FALSE POSITIVE COMPLAINTS. THIS APPROACH INVOLVES REVIEW OF MANUFACTURING BATCH HISTORY RECORDS, TESTING OF RETENTION SAMPLES, AND TESTING OF CUSTOMER RETURNED SAMPLES, IF APPLICABLE. AN INVESTIGATION AND TESTING WERE PERFORMED ON THE BATCH NUMBER PROVIDED. A PHOTO WAS PROVIDED AND SHOWED ¿COV2: +¿ IN THE DISPLAY SCREEN. THE REPORTED WAS UNABLE TO BE CONFIRMED. THE ROOT CAUSE COULD NOT BE IDENTIFIED. BD QUALITY WILL CONTINUE TO CLOSELY MONITOR FOR TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING PCR AND THE RESULT WAS NEGATIVE. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "REPORTING COMPLAINT AS FALSE POSITIVE RESULT."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WHILE TESTING FOR SARS COV-2 A FALSE POSITIVE RESULT WAS OBTAINED. A REPEAT TEST WAS PERFORMED USING PCR AND THE RESULT WAS NEGATIVE. THERE WAS NO INDICATION THAT RESULTS WERE REPORTED OUT AND THERE WAS NO REPORT OF PATIENT IMPACT. EUA#: (B)(4). THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "REPORTING COMPLAINT AS FALSE POSITIVE RESULT."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1269895 BD RAPID DETECTION OF SARS-COV-2 VERITOR CORONAVIRUS ANTIGEN DETECTION SYSTEM QKP BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. 1181838

Patients

Seq Age Sex Outcome Treatment
1