BINAX NOW COVID-19 AG CARD
Report
- Report Number
- 1221359-2021-02425
- Event Type
- Malfunction
- Date Received
- August 26, 2021
- Date of Event
- August 1, 2021
- Report Date
- December 21, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA202537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 195-160/LOT 152369 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-430 WL/LOT 148095. TEST BASE PART NUMBER 190-100J/LOT 152369. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 152369 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, THE MANUFACTURING BATCH RECORD REVIEW (BRR) REVEALED THAT THE PRODUCT MET ACCEPTANCE CRITERIA FOR RELEASE, AND REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES THAT THE PRODUCT IS PERFORMING ACCORDING TO THE STATEMENTS CONTAINED IN THE PACKAGE INSERT AND A PRODUCT DEFICIENCY HAS NOT BEEN IDENTIFIED. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE OR THE SPECIFIC PATIENT SAMPLE.
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CUSTOMER REPORTED AN UNCONFIRMED FALSE POSITIVE RESULT WITH THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2021 ON A TESTED NASAL KITTED SWAB AND GENERATED A POSITIVE RESULTS. THE CUSTOMER STATED HE WAS SYMPTOMATIC 12 HOURS AFTER RECEIVING THE PFIZER VACCINE ON (B)(6) 2021. THE CUSTOMER IS TAKING MULTIVITAMINS AND (B)(6). THE CUSTOMER IS ALSO SELF-ISOLATING. TECHNICAL SERVICES DID RECOMMEND THE CUSTOMER TO UNDERGO CONFIRMATORY TESTING AND WORKING WITH CUSTOMER'S PERSONAL PHYSICIAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1268017 | BINAX NOW COVID-19 AG CARD | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 152369 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Male |