FDA Adverse Event Malfunction Summary report: N

BINAX NOW COVID-19 AG CARD

MDR report key: 12369006 · Received August 26, 2021

Report

Report Number
1221359-2021-02425
Event Type
Malfunction
Date Received
August 26, 2021
Date of Event
August 1, 2021
Report Date
December 21, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA202537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 195-160/LOT 152369 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-430 WL/LOT 148095. TEST BASE PART NUMBER 190-100J/LOT 152369. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 152369 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, THE MANUFACTURING BATCH RECORD REVIEW (BRR) REVEALED THAT THE PRODUCT MET ACCEPTANCE CRITERIA FOR RELEASE, AND REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES THAT THE PRODUCT IS PERFORMING ACCORDING TO THE STATEMENTS CONTAINED IN THE PACKAGE INSERT AND A PRODUCT DEFICIENCY HAS NOT BEEN IDENTIFIED. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE OR THE SPECIFIC PATIENT SAMPLE.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN UNCONFIRMED FALSE POSITIVE RESULT WITH THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2021 ON A TESTED NASAL KITTED SWAB AND GENERATED A POSITIVE RESULTS. THE CUSTOMER STATED HE WAS SYMPTOMATIC 12 HOURS AFTER RECEIVING THE PFIZER VACCINE ON (B)(6) 2021. THE CUSTOMER IS TAKING MULTIVITAMINS AND (B)(6). THE CUSTOMER IS ALSO SELF-ISOLATING. TECHNICAL SERVICES DID RECOMMEND THE CUSTOMER TO UNDERGO CONFIRMATORY TESTING AND WORKING WITH CUSTOMER'S PERSONAL PHYSICIAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1268017 BINAX NOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 152369 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 42 YR Male