FDA Adverse Event Injury Summary report: N

LIGHT ADJUSTABLE LENS (LAL)

MDR report key: 12368535 · Received August 25, 2021

Report

Report Number
3012712027-2021-00046
Event Type
Injury
Date Received
August 25, 2021
Report Date
August 25, 2021
Manufacturer
RXSIGHT, INC.
Product Code
PZK
UDI-DI
00818806020166
PMA / PMN Number
P160055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

SITE REPORTED BILATERAL PATIENT INADVERTENTLY EXPOSED THE IMPLANTED LIGHT ADJUSTABLE LENSES TO AMBIENT UV LIGHT. BOTH OF THE LENSES WERE EXPLANTED. RXSIGHT'S AWARENESS DATE WAS 8/2/2021. THE EXPLANTED LENSES WERE RETURNED FOR EVALUATION. VISUAL INSPECTION AND OPTICAL TESTING OF BOTH RETURNED LENSES CONFIRMED THE PRESENCE OF AMBIENT ZONES ON THE LENSES, WHICH IS INDICATIVE OF EXPOSURE OF THE LENSES TO AMBIENT UV LIGHT PRIOR TO LOCK IN.

Description of Event or Problem · 0

SITE REPORTED BILATERAL PATIENT INADVERTENTLY EXPOSED THE IMPLANTED LIGHT ADJUSTABLE LENSES TO AMBIENT UV LIGHT. BOTH OF THE LENSES WERE EXPLANTED. RXSIGHT'S AWARENESS DATE WAS 8/2/2021.

Additional Manufacturer Narrative · 1

SITE REPORTED BILATERAL PATIENT INADVERTENTLY EXPOSED THE IMPLANTED LIGHT ADJUSTABLE LENSES TO AMBIENT UV LIGHT. BOTH OF THE LENSES WERE EXPLANTED. RXSIGHT'S AWARENESS DATE WAS 8/2/2021. THE DEVICE HISTORY RECORD FOR THE LENSES WERE REVIEWED. NO ISSUES WERE NOTED. SERIAL NUMBER: (B)(4), LOT NUMBER: L02-001937, IMPLANT DATE: UNKNOWN, MFG DATE: 11/20/2020, EXPIRATION DATE: 10/31/2023, EXPLANT DATE: UNKNOWN, UDI: (B)(4). SERIAL NUMBER: (B)(4), LOT NUMBER: L02-001415, IMPLANT DATE: UNKNOWN, MFG DATE: 11/16/2019, EXPIRATION DATE: 10/31/2022, EXPLANT DATE: UNKNOWN, UDI: (B)(4). THE LENSES HAVE BEEN RECEIVED AND ARE CURRENTLY BEING EVALUATED.

Description of Event or Problem · 1

SITE REPORTED BILATERAL PATIENT INADVERTENTLY EXPOSED THE IMPLANTED LIGHT ADJUSTABLE LENSES TO AMBIENT UV LIGHT. BOTH OF THE LENSES WERE EXPLANTED. RXSIGHT'S AWARENESS DATE WAS 8/2/2021.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1265139 LIGHT ADJUSTABLE LENS (LAL) LIGHT ADJUSTABLE LENS PZK RXSIGHT, INC. 60005 L02-001937 00818806020166

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention