FDA Adverse Event Malfunction Summary report: N

DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 12367606 · Received August 25, 2021

Report

Report Number
3004753838-2021-146670
Event Type
Malfunction
Date Received
August 25, 2021
Date of Event
July 31, 2021
Report Date
October 3, 2021
Manufacturer
DEXCOM, INC.
Product Code
QBJ
PMA / PMN Number
DEN170088
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

COM-2012070-21.

Additional Manufacturer Narrative · 0

COM-2012070-21.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. VOLTAGE MEASUREMENT: PASSED. PAIRING TEST/DOWNLOAD: FAILED. A REVIEW OF THE SHARE LOG WAS PERFORMED, AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS A DEFECTIVE TRANSMITTER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED. SUBSEQUENT TO THE INITIAL MDR, ADDITIONAL INFORMATION IS AVAILABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. A REVIEW OF THE SHARE LOGS WAS PERFORMED AND SIGNAL LOSS WAS FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE THE MOBILE DEVICE LOST POWER. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1262016 DEXCOM G6 CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR QBJ DEXCOM, INC. MT25042-4 2638

Patients

Seq Age Sex Outcome Treatment
1 39 YR