CHLORAPREP UNKNOWN
Report
- Report Number
- 3004932373-2021-00415
- Event Type
- Injury
- Date Received
- August 25, 2021
- Date of Event
- August 12, 2021
- Report Date
- November 2, 2021
- Manufacturer
- CAREFUSION 213, LLC 0113
- Product Code
- OJU
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- PHYSICIAN
Narratives
SAMPLES WERE AVAILABLE FOR EVALUATION. SAMPLES WERE RECEIVED ACTIVATED WITH NO SOLUTION PRESENT IN THE AMPOULE. NO POSSIBLE ROOT CAUSE COULD BE DETERMINED BASED ON VISUAL EXAMINATION OF THESE SAMPLES RETURNED NOR COULD THE ADVERSE REACTION BE VERIFIED. RETENTION SAMPLES FOR PN: 931490, LN: 0140871 WERE VISUALLY EXAMINED AND THEY DID NOT PRESENT ANY DISCOLORATION THAT WOULD INDICATE THE PRESENCE OF MICROBIAL CONTAMINATION. IN ACCORDANCE WITH INTERNAL PROCEDURES, THE AMPOULES AND SOLUTION USED TO MANUFACTURE PN: 931490, LN: 010871 WERE SUBMITTED TO CHEMISTRY AND MICROBIAL TESTING AT THE TIME THE PRODUCTS WERE MANUFACTURED AND STERILIZED. CHEMISTRY AND MICROBIOLOGY TEST RESULTS FOR BOTH THE AMPOULE LOT AND SOLUTION LOT WERE WITHIN SPECIFICATION ENSURING CHG/IPA ASSAY, IMPURITY LEVELS AND SOLUTION STERILITY WERE WITHIN SPECIFICATIONS. IN ADDITION, THE CHG RAW MATERIAL USED IN THE SOLUTION WAS ALSO SUBMITTED TO CHEMISTRY TESTING PRIOR TO BEING USED IN MANUFACTURING AND TEST RESULTS ALSO SHOW THAT IT WAS WITHIN SPECIFICATION. THE BATCH RECORD FOR PN: 931490, LN: 0140871 WAS REVIEWED AND THERE WAS A NON-CONFORMANCE FOUND DURING THE ASSEMBLY OF THE LOT DURING ONE OF THE ROUTINE QA INSPECTIONS. THE NON-CONFORMANCE FOUND WAS RELATED TO FOREIGN MATERIAL. THE LOT WAS NOT REJECTED IN ACCORDANCE WITH OUR SAMPLING PLAN. THE LOT HAD A QUALITY NOTIFICATION OPENED DUE TO THE SUPPLIER ISSUE. THE SUPPLIER COMMUNICATED THAT THE SOLUTION LOT USED TO MANUFACTURE PN: 9301490, LN: 0140871 WAS POTENTIALLY AFFECTED BY A MICROBIAL CONTAMINATION. THE PRODUCT WAS PLACED ON HOLD AND A SITUATIONAL ANALYSIS WAS PERFORMED TO EVALUATE THE RISK OF THE POTENTIAL CONTAMINATION. THE RESULTS OF THE INVESTIGATION INDICATED THAT THE POSSIBILITY OF THE CONTAMINANT IS NONE DUE TO THE ABILITY OF IPA (ONE OF CHLORAPREP¿S INGREDIENTS) TO KILL BACTERIA THROUGH THE COAGULATION AND/OR DENATURING OF PROTEINS AND MACROMOLECULES, WHICH LEADS TO DISRUPTION OF CYTOPLASMIC INTEGRITY, CELL LYSIS, AND INTERFERENCE WITH CELLULAR METABOLISM. DUE TO THE NATURE OF THE INGREDIENTS IN CHLORAPREP, THE POTENTIAL FOR INFECTIOUS COMPLICATIONS DUE TO A POSSIBLE MICROBIAL CONTAMINATION IS NON-EXISTENT. THE DESIGN FAILURE MODE AND EFFECTS ANALYSIS WAS REVIEWED, AND THE RISK OF A SKIN REACTION IS CONSIDERED. POSSIBLE POTENTIAL CAUSES INCLUDE INCREASED CHG CONCENTRATION, INCREASED IPA CONCENTRATION, CHEMICAL SPECIFICATIONS DO NOT MEET REQUIREMENTS, INCREASED IMPURITY OR LEACHABLE CONCENTRATIONS. DUE TO THE NATURE OF CHLORAPREP INGREDIENTS IT MAY CAUSE ALLERGIC REACTIONS. INSTRUCTIONS FOR USE INCLUDE AN ALLERGY ALERT. THE ALERT READS ¿THIS PRODUCT MAY CAUSE A SEVERE ALLERGIC REACTION. SYMPTOMS MAY INCLUDE WHEEZING/DIFFICULTY BREATHING, SHOCK, FACIAL SWELLING, HIVES, RASH. IF AN ALLERGIC REACTION OCCURS, STOP USE AND SEEK MEDICAL HELP RIGHT AWAY. DO NOT USE ON PATIENTS ALLERGIC TO CHLORHEXIDINE GLUCONATE OR ANY OTHER INGREDIENT IN THIS PRODUCT¿. THERE WERE NO FINDINGS IN THE INVESTIGATION OF THE MANUFACTURING AND RELEASE PROCESSES OF PN: 931490, LN: 0140871 THAT WOULD SUGGEST A CAUSE FOR AN ADVERSE REACTION. NO FURTHER ACTIONS ARE REQUIRED. BD WILL CONTINUE TO TRACK AND TREND.
IT WAS REPORTED BY THE FACILITY REPRESENTATIVE THAT THERE WERE ADVERSE REACTIONS TO CHLORAPREP. PER SUMMARY REPORT: SUSPECTED ISSUE WITH CHLORAPREP 1.5 ML -LOT#: 0140871 FEBRUARY 25TH MM RECEIVED 4 COMPLAINTS REGARDING CHLORAPREP APPLICATOR. CMN ALSO REPORTED AN INCREASED NUMBER OF ADVERSE REACTIONS 2021- N=14. THIS LOT MAY HAVE BEEN INTRODUCED IN DECEMBER (TO BE CONFIRMED), IF SO, TOTAL DEFECTS AFTER THE FIRST DELIVERY IS N=16. THERE IS AN OVERALL INCREASE IN ARS SINCE NOVEMBER 2020.
(B)(4) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. (B)(4).
IT WAS REPORTED BY THE FACILITY REPRESENTATIVE THAT THERE WERE ADVERSE REACTIONS TO CHLORAPREP. PER SUMMARY REPORT: SUSPECTED ISSUE WITH CHLORAPREP 1.5 ML -LOT # 0140871. FEBRUARY 25TH MM RECEIVED 4 COMPLAINTS REGARDING CHLORAPREP APPLICATOR. CMN ALSO REPORTED AN INCREASED NUMBER OF ADVERSE REACTIONS 2021- N=14. THIS LOT MAY HAVE BEEN INTRODUCED IN DECEMBER (TO BE CONFIRMED), IF SO, TOTAL DEFECTS AFTER THE FIRST DELIVERY IS N=16. THERE IS AN OVERALL INCREASE IN ARS SINCE NOVEMBER 2020.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1263393 | CHLORAPREP UNKNOWN | 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL | OJU | CAREFUSION 213, LLC 0113 | 0140871 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |