FDA Adverse Event Malfunction Summary report: N

VISUAL-ICE CRYOABLATION SYSTEM CHINA

MDR report key: 12360227 · Received August 24, 2021

Report

Report Number
2134265-2021-10562
Event Type
Malfunction
Date Received
August 24, 2021
Date of Event
July 30, 2021
Report Date
October 25, 2021
Manufacturer
GALIL MEDICAL INC.
Product Code
GEH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATED BY MANUFACTURER: DEVICE WAS NOT RETURNED FOR ANALYSIS. FIELD REPAIR BEGUN 11AUG2021 AND IS STILL ONGOING. THE PART NEEDED TO COMPLETE THE REPAIR HAS NOT BEEN RECEIVED. THE FIELD AGENT WAS ABLE TO CHECK THE CONSOLE LOG AND REPORTED AN ERROR: '80 - DIGI VOLTAGE POWER CHECK FAILED'. THE WORK ORDER IS STILL OPEN FOR COMPLETION AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT INSUFFICIENT ICE WAS CREATED BY THE CONSOLE. A VISUAL-ICE CRYOABLATION SYSTEM WAS SELECTED FOR USE DURING A CRYOABLATION PROCEDURE. DURING DEVICE PREPARATION, THE DEVICE COULD NOT BE "REFRIGERATED". THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE PATIENT IS STABLE WITH NO COMPLICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT INSUFFICIENT ICE WAS CREATED BY THE CONSOLE. A VISUAL-ICE CRYOABLATION SYSTEM WAS SELECTED FOR USE DURING A CRYOABLATION PROCEDURE. DURING DEVICE PREPARATION, THE DEVICE COULD NOT BE "REFRIGERATED". THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE PATIENT IS STABLE WITH NO COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1259854 VISUAL-ICE CRYOABLATION SYSTEM CHINA UNIT, CRYOSURGICAL, ACCESSORIES GEH GALIL MEDICAL INC. FPRCH6000-03

Patients

Seq Age Sex Outcome Treatment
1