FDA Adverse Event Other Summary report: N

COUCHMOUNT VARIAN EXACT

MDR report key: 1235673 · Received November 12, 2008

Report

Report Number
8043933-2008-00006
Event Type
Other
Date Received
November 12, 2008
Date of Event
October 13, 2008
Report Date
November 11, 2008
Manufacturer
BRAINLAB AG
Product Code
MUJ
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE HAS BEEN NO PT INJURY; HOWEVER, A RISK TO PT HEALTH COULD NOT BE EXCLUDED. SUMMARY OF EVAL: THE BRAINLAB DEVICE (INCLUDING THE LOCKING MECHANISM) WORKS AS INTENDED. THERE IS NO UNANTICIPATED RISK WITH THE BRAINLAB DEVICE DESIGN RELATED TO THIS OCCURRENCE. THE SITUATION OCCURRED DUE TO AN ERROR OF USE SINCE ACCORDING USER INSTRUCTION AS CONTAINED IN THE USER MANUAL HAVE NOT BEEN FULLY APPLIED IN THIS CASE (VERIFY THAT THE COUCHMOUNT IS SECURELY LOCKED EVERY TIME THE COUCHMOUNT IS ATTACHED). CORRECTIVE AND PREVENTIVE ACTION: THE USER HAS BEEN REMINDED ABOUT THE ACCORDING INSTRUCTIONS AND WARNINGS IN THE USER MANUAL REGARDING THE USE OF THE COUCHMOUNT.

Description of Event or Problem · 1

THE HOSP REPORTED THAT THE BRAINLAB COUCHMOUNT WAS DISPLACED FROM ITS POSITION, DUE TO A COLLISION WITH THE LINEAR ACCELERATOR. NO PT INJURY OCCURRED ACCORDING TO THE HOSP, HOWEVER, A RISK TO A PT HEALTH COULD NOT BE EXCLUDED, SINCE THE COUCHMOUNT IS USED TO FIXATE THE PT'S HEAD TO THE TREATMENT COUCH OF THE LINEAR ACCELERATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COUCHMOUNT VARIAN EXACT COUCHMOUNT VARIAN EXACT MUJ BRAINLAB AG 40100-0B NA

Patients

Seq Age Sex Outcome Treatment
1