FDA Adverse Event
Malfunction
Summary report: N
SFX 90-90 TABLE
MDR report key: 12355
·
Received March 29, 1994
Report
- Report Number
- MW1001278
- Event Type
- Malfunction
- Date Received
- March 29, 1994
- Date of Event
- March 27, 1994
- Report Date
- March 28, 1994
- Manufacturer
- GENERAL ELECTRIC CO.
- Product Code
- KXJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
FILMS WERE COMPLETED ON THE BACK AND HIP OF A PT AND A STRETCHER WAS PLACED NEXT TO THE TABLE FOR TRANSFER. AS THE TECH WENT TO THE OPPOSTIE SIDE OF THE STRETCHER, THE X-RAY TABLE BEGAN TO ELEVATE. THE TECHNICIAN IMMEDIATELY PULLED THE STRETCHER AWAY FROM THE TABLE THINKING THAT THE TABLE WOULD STOP. THE TECHNICIAN THEN INFORMED THE PT THAT HE WAS GOING TO BE IN THE ERECT POSITION. THE TABLE DID NOT STOP ANGULATING OR ELEVATING. THE PT WAS TOLD TO BEND DOWN AND WAS REMOVED FROM THE TABLE. THE TECHNICIAN THEN PUSHED THE EMERGENCY BUTTON. THE RPTR STATED THAT THE TOP OF THE TABLE WENT TWO FEET INTO THE 11 FOOT CEILING. ALL INTERLOCKS AND SAFETY MECHANISMS FAILED TO STOP THE TABLE. THE PT WAS NOT INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SFX 90-90 TABLE | X-RAY TABLE | KXJ | GENERAL ELECTRIC CO. | 46-262751G1/G2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |