ADVANTA 2 FRAME
Report
- Report Number
- 3006697241-2021-00071
- Event Type
- Malfunction
- Date Received
- August 20, 2021
- Date of Event
- July 29, 2021
- Report Date
- July 29, 2021
- Manufacturer
- HILL-ROM MEXICO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
THE HILLROM TECHNICIAN FOUND THE SIDERAIL CABLE ASSY NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL THE ADVANTA" 2 SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. TEST EACH OF THE BUTTONS TO CHECK THAT THEY ACTIVATE THE CORRECT FUNCTION AND THAT THEY DO NOT WORK INTERMITTENTLY BY PRESSING EACH BUTTON FOR SEVERAL SECONDS. EACH MOVEMENT MUST BE CONTINUOUS. REPLACE THE PENDANT IF NECESSARY. A SEARCH OF THE HILLROM MAINTENANCE RECORDS SHOWED HILLROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN OCTOBER 6TH 2020. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN HAS ORDER AND WILL REPLACED THE SIDERAIL CABLE (PART#141806) TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE HEAD OF THE BED WAS MOVING ON IT'S OWN. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1245893 | ADVANTA 2 FRAME | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | HILL-ROM MEXICO | P1190A000072 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |