FDA Adverse Event Malfunction Summary report: N

BINAX NOW COVID-19 AG CARD

MDR report key: 12346558 · Received August 20, 2021

Report

Report Number
1221359-2021-02333
Event Type
Malfunction
Date Received
August 20, 2021
Report Date
March 4, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
00811877011408
PMA / PMN Number
EUA202537
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 155512 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-160 / LOT 155512 AND TEST BASE PART NUMBER 195-430H / LOT 152829. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 155512 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN ADDITION, A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 155512 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, THE MANUFACTURING BATCH RECORD REVIEW (BRR) REVEALED THAT THE PRODUCT MET ACCEPTANCE CRITERIA FOR RELEASE, AND REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES THAT THE PRODUCT IS PERFORMING ACCORDING TO THE STATEMENTS CONTAINED IN THE PACKAGE INSERT AND A PRODUCT DEFICIENCY HAS NOT BEEN IDENTIFIED. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SELF-TEST USER PERFORMANCE, INTERPRETATION OF THE RESULT, OR THE SPECIFIC PATIENT SAMPLE.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED CONFLICTING RESULTS WITH THE BINAXNOW COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2021 ON AN UNKNOWN SAMPLE. THE CUSTOMER REPORTED THE FIRST BINAXNOW COVID-19 ANTIGEN SELF TEST GENERATED POSITIVE RESULTS. REPEAT TESTING WAS PERFORMED (B)(6) 2021 WITH BINAXNOW COVID-19 ANTIGEN SELF TEST AND GENERATED NEGATIVE RESULTS. PCR CONFIRMATION TESTING WAS NOT PERFORMED AT THE TIME OF THE CUSTOMER'S CALL. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1245368 BINAX NOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 155512 00811877011408

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other