ALTEON CUP, CLSTR HOLE G6, SIZE 54MM
Report
- Report Number
- 1038671-2021-00399
- Event Type
- Injury
- Date Received
- August 19, 2021
- Date of Event
- August 2, 2021
- Report Date
- July 15, 2022
- Manufacturer
- EXACTECH, INC.
- Product Code
- JDI
- UDI-DI
- 10885862550002
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
AFTER FURTHER REVIEW OF ADDITIONAL INFORMATION RECEIVED THE FOLLOWING SECTIONS G3, G7, H2 AND H6 HAVE BEEN UPDATED ACCORDINGLY. AS REPORTED, APPROXIMATELY SIX MONTHS POSTOP THE INITIAL LTHA, THIS 79 Y/O MALE PATIENT WAS REVISED DUE TO A PERI PROSTHETIC FRACTURE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION DUE TO HOSPITAL POLICY. THESE DEVICES WERE USED FOR TREATMENT, NOT DIAGNOSIS. A KNOWN SURGICAL RISK FOR TOTAL HIP REVISION IS POTENTIAL BONE FRACTURE. AS PART OF THE PREOPERATIVE ASSESSMENT, THE SURGEON MUST ENSURE THAT NO BIOLOGICAL, BIOMECHANICAL, OR OTHER FACTORS EXIST THAT MIGHT ADVERSELY AFFECT THE SURGERY AND/OR POSTOPERATIVE PERIOD. THE MOST LIKELY CAUSE FOR THE BONE FRACTURE POSSIBLY SURGICAL TECHNIQUE AND PATIENT CONDITIONS AT THE TIME OF IMPLANTATION OF THE TOTAL HIP COMPONENTS.
CONCOMITANT DEVICES: BIOLOX DELTA FEMORAL HEAD 40MM OD, +7MM (CAT# 170-40-07). CUP (CAT# 01-030-10-0654 / SERIAL# (B)(4)). ADDITIONAL INFORMATION, INCLUDING THE PRODUCT INVESTIGATION, WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.
AS REPORTED, APPROXIMATELY SIX MONTHS POSTOP THE INITIAL LTHA, THIS (B)(6) MALE PATIENT WAS REVISED DUE TO A PERI PROSTHETIC FRACTURE. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION DUE TO HOSPITAL POLICY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1238151 | ALTEON CUP, CLSTR HOLE G6, SIZE 54MM | PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED | JDI | EXACTECH, INC. | ALT XLE LNR EXTCOV G6 40 | UNK | 10885862550002 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Male | Required Intervention |