FDA Adverse Event Malfunction Summary report: N

COLLEAGUE 3 VOLUMETRIC INFUSION PUMP

MDR report key: 1234104 · Received November 19, 2008

Report

Report Number
6000001-2007-03301
Event Type
Malfunction
Date Received
November 19, 2008
Date of Event
January 1, 2007
Report Date
January 9, 2007
Manufacturer
BAXTER HEALTHCARE PTE. LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN REQUESTED BY BAXTER FOR EVALUATION, BUT HAS NOT YET BEEN RECEIVED. SHOULD THE PUMP BE RECEIVED FOR EVALUATION, A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: DEVICE EVALUATED IN 2007. FAILURE CODE 570 WAS CONFIRMED. FAILURE CODE 570 IS CAUSED BY LOW BATTERY VOLTAGE. DURING INSPECTION THE DEVICE FAILED THE BATTERY DISCHARGE TEST WHICH WAS CAUSED BY DEPLETED BATTERIES THAT WERE POTENTIALLY DAMAGED AND THEREFORE REPLACED.

Description of Event or Problem · 1

THE FACILITY REPORTED AN INFUSION PUMP WITH FAILURE CODE 570. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE DEVICE WAS IN PATIENT USE WHEN THE EVENT OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE 3 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE PTE. LTD. N/A NA