COLLEAGUE 3 VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-03301
- Event Type
- Malfunction
- Date Received
- November 19, 2008
- Date of Event
- January 1, 2007
- Report Date
- January 9, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS BEEN REQUESTED BY BAXTER FOR EVALUATION, BUT HAS NOT YET BEEN RECEIVED. SHOULD THE PUMP BE RECEIVED FOR EVALUATION, A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
EVALUATION SUMMARY: DEVICE EVALUATED IN 2007. FAILURE CODE 570 WAS CONFIRMED. FAILURE CODE 570 IS CAUSED BY LOW BATTERY VOLTAGE. DURING INSPECTION THE DEVICE FAILED THE BATTERY DISCHARGE TEST WHICH WAS CAUSED BY DEPLETED BATTERIES THAT WERE POTENTIALLY DAMAGED AND THEREFORE REPLACED.
THE FACILITY REPORTED AN INFUSION PUMP WITH FAILURE CODE 570. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE DEVICE WAS IN PATIENT USE WHEN THE EVENT OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.