FDA Adverse Event Malfunction Summary report: N

COBAS 6000 C501 MODULE

MDR report key: 1234084 · Received November 4, 2008

Report

Report Number
1823260-2008-08144
Event Type
Malfunction
Date Received
November 4, 2008
Date of Event
October 24, 2008
Report Date
November 4, 2008
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CEM
PMA / PMN Number
K060373
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Description of Event or Problem · 1

USER EXPERIENCED A LEAK COMING FROM THE WATER TANK OF THE ANALYZER DUE TO A BROKEN WATER TANK VALVE. NO ONE FELL OR WAS INJURED, AND NO PT SAMPLES WERE INVOLVED. THE FIELD SERVICE REP FOUND A VALVE BROKEN OFF OF THE WATER TANK, AND REPLACED COMPLETE WATER TANK ASSEMBLY. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPEC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COBAS 6000 C501 MODULE CLINICAL CHEMISTRY ANALYZER CEM ROCHE DIAGNOSTICS C501

Patients

Seq Age Sex Outcome Treatment
1 UNK