FDA Adverse Event
Malfunction
Summary report: N
COBAS 6000 C501 MODULE
MDR report key: 1234084
·
Received November 4, 2008
Report
- Report Number
- 1823260-2008-08144
- Event Type
- Malfunction
- Date Received
- November 4, 2008
- Date of Event
- October 24, 2008
- Report Date
- November 4, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- CEM
- PMA / PMN Number
- K060373
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
USER EXPERIENCED A LEAK COMING FROM THE WATER TANK OF THE ANALYZER DUE TO A BROKEN WATER TANK VALVE. NO ONE FELL OR WAS INJURED, AND NO PT SAMPLES WERE INVOLVED. THE FIELD SERVICE REP FOUND A VALVE BROKEN OFF OF THE WATER TANK, AND REPLACED COMPLETE WATER TANK ASSEMBLY. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPEC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS 6000 C501 MODULE | CLINICAL CHEMISTRY ANALYZER | CEM | ROCHE DIAGNOSTICS | C501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |