FDA Adverse Event
Injury
Summary report: N
GENTLEPOWER LUX CONTRA ANGLE 25LPA
MDR report key: 1233984
·
Received November 19, 2008
Report
- Report Number
- 8010493-2008-00022
- Event Type
- Injury
- Date Received
- November 19, 2008
- Report Date
- October 22, 2008
- Manufacturer
- KALTENBACH & VOIGT GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE PATIENT WAS PRESCRIBED SILVER SULFADIAZINE TO APPLY TO BURN ON LIP. THE PATIENT COMPLETELY HEALED. AT THE HANDPIECE EVALUATION, IT WAS FOUND THAT THE HEAD WAS DENTED, THE BEARINGS WERE GRINDING, AND THE HEAD WAS HEATING UP. LOW RESIDUE LEVEL WAS PRESENT. THE OFFICE WAS INFORMED THAT THE DENT IN THE HEAD WAS CAUSING IT TO OVERHEAT. THE HANDPIECE WAS REPAIRED.
Description of Event or Problem · 1
WHILE THE DENTIST WAS WORKING ON TOOTH #22 AND #23, THE PATIENT'S LIP WAS BURNED ABOUT THE SIZE OF A DIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GENTLEPOWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECE AND ACCESSORIES | EFB | KALTENBACH & VOIGT GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |