FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL PRECISE

MDR report key: 12339297 · Received August 18, 2021

Report

Report Number
8041187-2021-00741
Event Type
Malfunction
Date Received
August 18, 2021
Date of Event
June 22, 2021
Report Date
August 30, 2021
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY BD RECEIVED THE SAMPLE FROM THE CUSTOMER'S FACILITY FOR INVESTIGATION. THE SAMPLES WAS EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE OF FOREIGN MATTER WAS OBSERVED. BASED ON THE FTIR RESULTS, THE TYPE OF FOREIGN MATTER COULD NOT BE IDENTIFIED BUT IT SHOWED FEW PEAKS SIMILAR TO THAT OF SYRINGE MATERIAL, POLYPROPYLENE. PROBABLE CAUSE OF EMBEDDED FOREIGN MATTER COULD BE A RESULT OF DEGRADED POLYPROPYLENE IN AN AGING HYDRAULIC PRESS, CLEANING OF THE INJECTION SCREW OF THE MOLDING MACHINE. A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0233477 THE BATCH WAS MANUFACTURED ON SYRINGE ASSEMBLY 50ML IN SEP-2020. NO SIMILAR DEFECT FOUND DURING IN-PROCESS INSPECTION AND QA OUTGOING INSPECTION. NO QUALITY NOTIFICATION WAS RAISED ON PAST 12 MONTHS ON SIMILAR NON-CONFORMANCE. H3 OTHER TEXT : SEE H10

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 50ML LL PRECISE HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BROWN RESIDUE IN THE SYRINGE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 50ML LL PRECISE HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: BROWN RESIDUE IN THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1234381 SYRINGE 50ML LL PRECISE SYRINGE FMF BECTON DICKINSON MEDICAL (SINGAPORE) 0233477

Patients

Seq Age Sex Outcome Treatment
1