Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS FOLLOW UP REPORT, WILLIAM COOK EUROPE INFORMS THAT THIS COMPLAINT HAS BEEN TRANSFERRED FROM WILLIAM COOK EUROPE TO COOK INC. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED. BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE.

Description of Event or Problem · 0

ADDITIONAL INFORMATION PROVIDED DETERMINED THAT THIS DEVICE WAS MANUFACTURED BY COOK INC. WITH THE SUBMISSION OF THIS FOLLOW UP REPORT, WILLIAM COOK EUROPE INFORMS THAT THIS COMPLAINT HAS BEEN TRANSFERRED FROM WILLIAM COOK EUROPE TO COOK INC.

Additional Manufacturer Narrative · 1

INITIAL REPORTER OCCUPATION: NON-HEALTHCARE PROFESSIONAL. INVESTIGATION: THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: ORGAN (VERTEBRAL)/VC PERFORATION THE REPORTED ALLEGATIONS HAVE BEEN INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: TRAUMA TO ADJACENT STRUCTURES, VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. CATALOG NUMBER AND LOT NUMBER ARE UNKNOWN. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FOLLOWING INFORMATION IS ALLEGED: THE PATIENT RECEIVED A CELECT INFERIOR VENA CAVA (IVC) FILTER ON (B)(6) 2009. APPROXIMATELY 9 YEARS AND 7 MONTHS LATER, THE PATIENT UNDERWENT A STENT PLACEMENT RELATED TO A CARDIAC PROCEDURE, AT WHICH TIME, DOCTORS DISCUSSED THE FACT THAT THE FILTER WAS NO LONGER RETRIEVABLE. AROUND 1 MONTH LATER, THE PATIENT UNDERWENT A COMPUTED TOMOGRAPHY (CT) SCAN OF THE ABDOMEN WITHOUT CONTRAST, WHICH REVEALED FILTER STRUT PERFORATION INTO THE AORTA AS WELL AS HIS LUMBAR SPINE AT L3. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED, BUT NOT YET PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1236604 UNKNOWN DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening