BINAX NOW COVID-19 AG CARD
Report
- Report Number
- 1221359-2021-02316
- Event Type
- Malfunction
- Date Received
- August 18, 2021
- Report Date
- December 18, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- QKP
- UDI-DI
- 00811877011408
- PMA / PMN Number
- EUA202537
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE INVESTIGATION CONCLUSION. TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 195-160/ LOT 153476 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-430H / LOT 148405. TEST BASE PART NUMBER 190-100 / LOT 153476. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 153476 SHOWED THAT THE COMPLAINT RATE IS (B)(4). IN CONCLUSION, THE MANUFACTURING BATCH RECORD REVIEW (BRR) REVEALED THAT THE PRODUCT MET ACCEPTANCE CRITERIA FOR RELEASE, AND REVIEW OF COMPLAINTS AGAINST THE KIT LOT FOR THE REPORTED ISSUE INDICATES THAT THE PRODUCT IS PERFORMING ACCORDING TO THE STATEMENTS CONTAINED IN THE PACKAGE INSERT AND A PRODUCT DEFICIENCY HAS NOT BEEN IDENTIFIED. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD POSSIBLY BE RELATED TO ISSUES INCLUDING THE SPECIFIC PATIENT SAMPLE.
THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.
THE CUSTOMER REPORTED AN UNCONFIRMED FALSE POSITIVE RESULT WITH THE BINAXNOW¿ COVID-19 ANTIGEN SELF TEST PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED NASAL KITTED SWAB. REPEAT TESTING WAS NOT PERFORMED AS THE CONSUMER WANTED TO PERFORM THE SECOND TEST AT THEIR PHYSICIANS OFFICE. THE CUSTOMER REPORTED THAT SHE GOT A SOLID PINK CONTROL LINE AND SOLID PINK SAMPLE LINE, BUT SHE'S NOT SURE IF THIS IS ACCURATE AND VALID BECAUSE WHILE ROTATING THE NASAL SWAB FROM THE TOP HOLE, SOME OF THE REAGENT SOLUTION SPILLED OUT OF THE HOLE. THE CONSUMER DIDN'T KNOW IF SHE HAD A COLD BUT SHE WAS WORRIED BECAUSE 7 OR 10 DAYS AGO SHE WAS EXPOSED TO SOMEONE WHO WAS POSITIVE WITH COVID-19 AND THAT IS WHY SHE WAS PERFORMING THE TEST. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1235762 | BINAX NOW COVID-19 AG CARD | LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 | QKP | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 153476 | 00811877011408 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |