FDA Adverse Event Injury Summary report: N

SPECTRANETICS GLIDELIGHT LASER SHEATH

MDR report key: 12334665 · Received August 17, 2021

Report

Report Number
1721279-2021-00155
Event Type
Injury
Date Received
August 17, 2021
Date of Event
January 1, 2013
Report Date
July 20, 2021
Manufacturer
THE SPECTRANETICS CORPORATION
Product Code
MFA
PMA / PMN Number
P960042
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RELATED MDRS: 1721279-2021-00156, 1721279-2021-00157,1721279-2021-00158, 1721279-2021-00159. THE DEVICE WAS NOT RETURNED FOR EVALUATION, THUS NO INVESTIGATION COULD BE COMPLETED.

Description of Event or Problem · 1

A PHILIPS EMPLOYEE BECAME AWARE ON 20 JULY 2021 OF A JOURNAL ARTICLE TITLED "OUTCOMES OF TRANSVENOUS LEAD EXTRACTION USING THE TIGHTRAIL¿ MECHANICAL ROTATING DILATOR SHEATH AND EXCIMER LASER SHEATH" (WILEY, DOI: 10.1111/JCE.15105, ACCEPTED 22 APRIL 2021). IN THE ARTICLE, DATA WAS COLLECTED FROM THE TRANSVENOUS LEAD EXTRACTION (TLE) REGISTRY AT ATRIUM HEALTH CAROLINAS MEDICAL CENTER. IT WAS DOCUMENTED THAT FIVE VASCULAR LACERATIONS AND CARDIAC AVULSIONS OCCURRED AND SPECTRANETICS DEVICES WERE IN USE (2 GLIDELIGHT OR SLS, 2 TIGHTRAIL, AND 1 GLIDELIGHT/SLS OR TIGHTRAIL). BECAUSE THE ARTICLE DID NOT A SPECIFY PROCEDURE DATE FOR THE ADVERSE EVENT DESCRIBED, THE PROCEDURE DATE OF 01/01/2013 HAS BEEN POPULATED SINCE THE STUDY INCLUDED DATA FROM 2013-2019. CASE DETAIL, LEAD INFORMATION, INTERVENTION SPECIFICS (OTHER THAN STERNOTOMY) AND PATIENT OUTCOME CANNOT BE DETERMINED FROM THE DOCUMENTATION AVAILABLE. IT CANNOT BE DETERMINED FROM THE ARTICLE WHETHER A SPECTRANETICS GLIDELIGHT OR SLS LASER SHEATH WAS UTILIZED, SO THE MANUFACTURER CHOSE A GLIDELIGHT DEVICE FOR THIS COMPLAINT/ADVERSE EVENT. ADDITIONAL INFORMATION IS BEING SOUGHT FROM THE AUTHOR. THIS COMPLAINT CAPTURES #1 OF 2 GLIDELIGHT DEVICES IN WHICH A VASCULAR LACERATION AND CARDIAC AVULSION OCCURRED AND REQUIRED INTERVENTION, INCLUDING STERNOTOMY. THERE WAS NO ALLEGED MALFUNCTION OF ANY SPECTRANETICS DEVICE IN USE DURING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1231884 SPECTRANETICS GLIDELIGHT LASER SHEATH DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS MFA THE SPECTRANETICS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention