FDA Adverse Event Malfunction Summary report: N

TUBE CIT PLH 13X75 2.7 109 USI L-MG CE

MDR report key: 12333548 · Received August 17, 2021

Report

Report Number
9617032-2021-00889
Event Type
Malfunction
Date Received
August 17, 2021
Date of Event
July 30, 2021
Report Date
August 19, 2021
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: BD DID RECEIVE 2 PHOTOGRAPHS FROM THE CUSTOMER IN SUPPORT OF THIS COMPLAINT, SHOWING ACCEPTABLE DRAW LEVEL. ADDITIONALLY, 20 RETAINED SAMPLES WERE FUNCTIONALLY TESTED AND THE INDICATED FAILURE MODE OF OVERFILL WAS NOT OBSERVED AS ALL RESULTS WERE WITHIN SPECIFICATION. BD WAS UNABLE TO CONFIRM CUSTOMERS INDICATED FAILURE MODE BASED ON THE RETAINED SAMPLES TEST RESULTS AND PHOTOGRAPHS ATTACHED. NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. THE DEVICE HISTORY RECORDS WERE REVIEWED WITH NO ISSUES BEING IDENTIFIED. THERE WERE NO RELATED QUALITY NOTIFICATIONS. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE TUBE CIT PLH 13X75 2.7 109 USI L-MG CE THERE WAS OVERFILL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE IS "OVERFILLING OF THE TUBE. NOT USABLE FOR ANALYSIS."

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED WHEN USING THE TUBE CIT PLH 13X75 2.7 109 USI L-MG CE THERE WAS OVERFILL. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: THERE IS "OVERFILLING OF THE TUBE. NOT USABLE FOR ANALYSIS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1227145 TUBE CIT PLH 13X75 2.7 109 USI L-MG CE BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 1020146

Patients

Seq Age Sex Outcome Treatment
1