FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1233203
·
Received November 18, 2008
Report
- Report Number
- 6000001-2007-03879
- Event Type
- Malfunction
- Date Received
- November 18, 2008
- Date of Event
- January 12, 2007
- Report Date
- January 12, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE REPORTED CONDITION OF DEPLETED BATTERIES WAS CONFIRMED. INSPECTION OF THE DEVICE FOUND THAT THE MAIN BATTERIES WERE POTENTIALLY DAMAGED AND WERE THEREFORE REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED.
Description of Event or Problem · 1
THE FACILITY RETURNED THE DEVICE FOR SERVICE. DURING PRODUCT EVALUATION, THE PUMP¿S BATTERIES WERE FOUND TO BE DEPLETED. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |