FDA Adverse Event
Malfunction
Summary report: N
COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1233199
·
Received November 18, 2008
Report
- Report Number
- 6000001-2007-02791
- Event Type
- Malfunction
- Date Received
- November 18, 2008
- Date of Event
- January 1, 2007
- Report Date
- January 3, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE LTD
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE CONDITION OF FAIL CODE 812:02 WAS CONFIRMED IN THE EVENT HISTORY. A FAULTY PUMP HEAD MODULE CAUSED THIS FAIL CODE. THE PUMP HEAD MODULE WAS REPLACED.
Description of Event or Problem · 1
THE FACILITY REPRESENTATIVE REPORTED FAILURE CODE 812:02. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE FAILURE OCCURRED DURING A PATIENT INFUSION. THE HOSPITAL REPRESENTATIVE STATED THAT THERE WERE NO REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE LTD | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |