COLLEAGUE CX VOLUMETRIC INFUSION PUMP

MDR report key: 1233182 · Received November 18, 2008

Report

Report Number
6000001-2007-03958
Event Type
Malfunction
Date Received
November 18, 2008
Date of Event
January 1, 2007
Report Date
January 2, 2007
Manufacturer
BAXTER HEALTHCARE PTE.LTD.
Product Code
FRN
Removal / Correction Number
CONTINUED ON H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: EVALUATION WAS PERFORMED AND THE REPORTED CONDITION OF BATTERY WILL NOT HOLD CHARGE WAS CONFIRMED DUE TO DEPLETED BATTERIES. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES CAUSED FAILURE CODE 570:320:844:0000. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPORTED AN INFUSION PUMP THAT WILL NOT HOLD BATTERY CHARGE. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE DEVICE WAS IN PATIENT USE WHEN THE EVEN OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.