COLLEAGUE CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-03958
- Event Type
- Malfunction
- Date Received
- November 18, 2008
- Date of Event
- January 1, 2007
- Report Date
- January 2, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE.LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: EVALUATION WAS PERFORMED AND THE REPORTED CONDITION OF BATTERY WILL NOT HOLD CHARGE WAS CONFIRMED DUE TO DEPLETED BATTERIES. INSPECTION OF THE PUMP REVEALED DEPLETED MAIN BATTERIES CAUSED FAILURE CODE 570:320:844:0000. THE MAIN BATTERIES WERE REPLACED. THE PUMP WAS TESTED FOR PROPER OPERATION AND PUMP FOUND TO FUNCTION PROPERLY. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS CURRENTLY BEING INVESTIGATED UNDER CAPA.
THE FACILITY REPORTED AN INFUSION PUMP THAT WILL NOT HOLD BATTERY CHARGE. INFORMATION WAS NOT PROVIDED REGARDING WHETHER OR NOT THE DEVICE WAS IN PATIENT USE WHEN THE EVEN OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE.LTD. | N/A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |