FDA Adverse Event Malfunction Summary report: N

FLO-GARD 6301 VOLUMETRIC INFUSION PUMP

MDR report key: 1233177 · Received November 18, 2008

Report

Report Number
6000001-2007-03314
Event Type
Malfunction
Date Received
November 18, 2008
Date of Event
January 1, 2007
Report Date
January 3, 2007
Manufacturer
BAXTER HEALTHCARE (SG)
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: A BAXTER SERVICE TECHNICIAN EVALUATED THE PUMP AND FOUND BROKEN DOOR HINGES. THE REPORTED CONDITION OF BROKEN DOOR #2 WAS CONFIRMED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED BROKEN DOOR #2 DURING CUSTOMER USE. ALTHOUGH BAXTER ATTEMPTED TO OBTAIN INFORMATION, DETAILS WERE NOT AVAILABLE REGARDING ADDITIONAL CONTACT INFORMATION. ACCORDING TO THE FACILITY REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLO-GARD 6301 VOLUMETRIC INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE (SG) NA NA

Patients

Seq Age Sex Outcome Treatment
1