COLLEAGUE CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-05964
- Event Type
- Malfunction
- Date Received
- November 18, 2008
- Date of Event
- April 1, 2007
- Report Date
- April 2, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 2005-047-MD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
FAILURE CODE 535:320:844:0000 WAS INITIALLY REPORTED BY THE FACILITY. IT IS UNKNOWN WHEN THE FAILURE CODE OCCURRED. EVALUATION SUMMARY: DURING PRODUCT EVALUATION, A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB) WAS OBSERVED. FAILURE CODE 535:320:844:0000 IN THE EVENT HISTORY CONFIRMS THE DEFECTIVE UIM PCB. THIS FAILURE CODE IS MANIFESTED AS A RESULT OF THE UIM PCB BEING DEFECTIVE. THE UIM PCB WAS REPLACED. REVIEW OF THE COMPLAINT HISTORY REVEALS SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT FOR THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED.
DURING SERVICE BY BAXTER, THE INFUSION PUMP WAS FOUND TO CONTAIN A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION OR CONTACT WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | N/A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |